LaMotte SMART3 Soil Operator's Manual page 52

Colorimeter
Hide thumbs Also See for SMART3 Soil:
Table of Contents

Advertisement

1.800.544.2843
12. Insert the low standard.
Press
ENTER
. The Insert
High Standard screen will be
displayed.
13. Insert the high standard. Press
ENTER
. The meter will display
the constants for the calibration
curve.
14. Press
ENTER
to save the
calibration curve and return to
the Edit User Test Menu.
SMART3 Soil Colorimeter 9.18
www.calcert.com
15
10
20
5
25
0
30
STD Calibration
Insert High Standard
<Enter> continue
12:00:00
001/500
K0 = 0.57246
K1 = 7.14547
K2 = 0.00000
K3 = 0.00000
<Enter> Save
12:00:00
001/500
111 H2O
Name the Test
Select Vial/WL
STD Calibration
Enter Constants
12:00:00
001/500
53