LaMotte SMART3 Soil Operator's Manual page 22

Colorimeter
Hide thumbs Also See for SMART3 Soil:
Table of Contents

Advertisement

1.800.544.2843
5. Press
to select the
ENTER
test.
6. Insert the blank into the
chamber. Close the lid.Press
to scan the blank. The
ENTER
screen wil display Scan Blank
Blank Done for about 1 second
and then return to the test
menu.
7.
Insert the reacted sample into
the chamber. Close the lid. Press
to scan the sample. The
ENTER
screen will display Scan Sample
Sample Done for about 1
second. The result will appear on
the screen.
8. To repeat the test, press
to scan the sample again. The
last blank scanned is used by
the colorimeter for repeated
scans. A diff erent blank can
be used by pressing
to scroll to Scan Blank
and then scanning another
blank. Scroll with
and make another
selection with
. The %T or
ENTER
Absorbance of the last test can
be viewed by scrolling down and
choosing %T/Abs. Press
to escape to previous menus.
NOTE: The menus loop in this
screen so either
will lead to the menu selection
needed.
SMART3 Soil Colorimeter 9.18
www.calcert.com
15
10
20
5
25
0
30
005 Aluminum
Scan Bank
Scan Sample
12:00:00
001/500
005 Aluminum
Scan Blank
Scan Sample
12:00:00
001/500
005 Aluminum
1.00
ppm
Scan Blank
Scan Sample
12:00:00
001/500
005 Aluminum
ENTER
1.00 ppm
Scan Bank
or
Scan Sample
12:00:00
001/500
or
EXIT
To Use the Print Test Function,
See the Bluetooth Menu Selection,
or
(page 63).