LaMotte SMART3 Soil Operator's Manual page 25

Colorimeter
Hide thumbs Also See for SMART3 Soil:
Table of Contents

Advertisement

6. Follow the test procedure in the
7.
8. The displayed result can now be
9. Press
26
1.800.544.2843
manual to test the prepared
standard. Insert the blank into
the chamber. Close the lid. Press
to scan the blank. The
ENTER
screen will display Scan Blank
Blank Done for about 1 second
and then return to the Test
Menu.
Insert the reacted standard
solution into the chamber. Close
the lid. Press
to scan the
ENTER
sample. The screen will display
Scan Sample Sample Done for
about 1 second. The result will
appear on the screen.
standardized. Press
or
to scroll to Calibrate.
to select Calibrate.
ENTER
A reverse font (light background
with dark characters) will appear
to indicate that the reading can
be adjusted.
www.calcert.com
15
10
20
5
25
0
30
005 Aluminum
Scan Blank
Scan Sample
12:00:00
001/500
005 Aluminum
0.28 ppm
Scan Blank
Scan Sample
12:00:00
001/500
005 Aluminum
0.28 ppm
%T/Abs
Calibrate
12:00:00
001/500
005 Aluminum
0.28 ppm
, =Edit, ENTER=Save
+ENTER=Default
12:00:00
001/500
SMART3 Soil Colorimeter 9.18