E.12 Incorrect Tray
Error: The correct tray is required.
Action: Open drawer, insert the correct tray and test device. Close drawer to
continue. If problem persists contact Abbott Technical Support.
E.13 Patient ID Invalid
Error: Patient ID Invalid.
Action: Patient ID must be 2 to 20 characters long, contain only alphanumeric
and some special characters, such as # () * /\ and have no spaces. Tap OK to try
again. If the error persists, contact Abbott Technical Support.
E.14 Test Device QR Code Read Failure
Error: Test device QR code read failure.
Action: Open drawer, check test device and orientation are correct, QR code
is not obscured and close drawer to continue. If error persists please contact
Abbott Technical Support.
E.15 Test Device ID Mismatch
Error: Foil pouch barcode and test device QR code do not match
Action: Open drawer, insert the test device from the scanned foil pouch and
close drawer to continue. Ensure the test device is from the foil pouch which
was scanned. If problem persists, contact Abbott Technical Support.
E.16 Test Device Present
Error: Test device is present in the drawer.
Action: Open drawer, remove the test device and close drawer to continue. If
the error persists, contact Abbott Technical Support.
E.17 Calibration Check Card Expired
Error: Calibration Check Card Expired.
Action: Open drawer, replace with an in-date Calibration Check Card and
close drawer to continue.
56
DIGIVAL User Manual