Caution:
Do NOT open the DIGIVAL drawer during testing. The test is not saved
until the completed result is displayed.
Never move the DIGIVAL while a test is in progress.
44
STEP 3: Verify that the test type
matches the QC sample intended for
testing and tap OK to confirm.
If the incorrect QC test has been
selected tap Back to return to the
previous screen.
STEP 4: Refer to specific test
Product Insert for QC preparation
and test procedure. Insert Test Device
at appropriate time according to Read
Mode chosen.
Read Now Mode: At the read time
indicated in assay instructions,
open drawer, insert test device,
and close drawer to continue.
Do NOT open drawer again until
result is displayed.
Walk Away Mode: Once the
test device is securely closed,
immediately open the drawer,
insert test device and close the
drawer to continue. The DIGIVAL
will time the test development and
read the result at the read time.
Do NOT open the drawer again
until the result is displayed.
STEP 5: Wait for the test completion.