Assays; Walk Away Mode - Abbott DIGIVAL LFR-024 User Manual

Hide thumbs Also See for DIGIVAL LFR-024:
Table of Contents

Advertisement

7.3.4 Assays

Home>Settings>Test Settings>Assays
Use the Assays menu option to configure new assays on the DIGIVAL, as they
become available.

7.3.5 Walk Away Mode

Home>Settings>Test Setting>Walk Away Mode
1. Tap Add Assay
2. Scan the 2D barcode provided with
the new assay kit.
3. The instrument will reboot itself
to complete the new assay
configuration.
Note: "Assay is already up-to-date" will
display on screen if attempt is made to
add an existing assay.
Use the Walk Away Mode menu
option to configure patient tests to be
run in Walk Away Mode. Walk Away
Mode allows the DIGIVAL to time
the test for the user, so that the read
occurs at the time specified in that
assay's product insert.
Note: Not all assays will have this
option. Refer to Product Insert for
details.
DIGIVAL User Manual
27