2. Set the date and time
To set the date and time, tap:
Settings>Set date and time
See Section 7.1, "Date and Time" for more detail.
3. Set the Admin password
To set the Admin password, tap:
Settings>Access Control>Admin password
Enter the new Admin password. See Section 7.2.4, "Set the Admin password" for
more detail.
4. Select operator login option
To select options, tap:
Settings>Access Control>Login
Select Operator ID only or Operator ID and Password. See section 7.2.1 for
more detail.
5. Create operator(s)
To create operator login permissions, tap:
Settings>Access Control>Operators>Actions>Add
Add at least one operator so that you can perform a Calibration Check. You can
return to this screen to add the rest of the staff at any time.
See Section 7.2.2.1, "Add a new Operator."
6. Log out as the admin
To log out, tap:
Home>Log Out
7. Log in as one of the newly created operators.
Scan or Enter Operator ID: Enter the operator login name and tap OK.
Password: If a password was specified in the Create operator screen, you will be
asked to enter your assigned password. Enter it and then tap OK.
8. Perform a Calibration Check
When the DIGIVAL is powered on for the first time it will require a calibration
check using the provided Calibration Check Card.
To start the check, tap:
Home>Calibration Check
Screen instructions guide you through the Calibration Check procedure. See
Section 6, "DIGIVAL Calibration" for more detail.