Do not read Test Results before or after their indicated read time, as the
results may not be correct.
If the DIGIVAL is configured to automatically upload test results, the result will
be uploaded to the configured middleware system at this point.
Caution:
Do NOT open the DIGIVAL drawer during testing. The test is not saved
until the completed result is displayed.
Never move the DIGIVAL while a test is in progress.
If the optional printer is set up, tap Print to print the test result. Tap Home to
return to the home screen or tap OK to continue.
If the DIGIVAL is set to Auto Print, the test result will automatically print.
Note: If the assay contains more than one test line, the individual results of each
test line will be reported.
Read Now Mode: At the read time
indicated in assay instructions,
open drawer, insert test device, and
close drawer to continue. Do NOT
open drawer again until result is
displayed.
Walk Away Mode: Once the
test device is securely closed,
immediately open the drawer, insert
test device and close the drawer to
continue. The DIGIVAL will time
the test development and read the
result at the read time. Do NOT
open the drawer again until the
result is displayed.
8. When a test completes
successfully, results are displayed
on the screen, with Positive or
Negative result. If a test error
occurs, 'Invalid' displays. Refer to
specific package insert for further
details on test result.
DIGIVAL User Manual