Guidance and manufacturer's declaration –
electromagnetic immunity
MAXFRAME AUTOSTRUT System is intended for use in the
electromagnetic environment specified below.
The customer or the user of the MAXFRAME AUTOSTRUT
System should assure that it is used in such an environment.
Declaration – electromagnetic immunity
Immunity test
IEC 60601
test level
Conducted RF
3 V, 6 V
IEC 61000-4-6
Radiated RF
3 V/m
IEC 61000-4-3
3 V from 0.15 to
80 MHz; 6 V from
0.15 to 80 MHz and
80% AM at 1 kHz
3 V/m from 80 MHz
to 2.7 GHz
Compliance
Electromagnetic environment – guidance
level
Portable and mobile RF communications equipment
should be used no closer to any part of the system,
including cables, than the recommended separation
distance calculated from the equation applicable to
the frequency of the transmitter.
Recommended separation distance
3 Vrms, 6 V
d = [
3 V/m
d = [
d = [
d = [
3 V from 0.15 to
Where P is the maximum output power rating of
80 MHz; 6 V from
the transmitter in watts (W) according to the trans-
0.15 to 80 MHz and
mitter manufacturer and d is the recommended
80% AM at 1 kHz
separation distance in metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.
3 V/m from 80 MHz
D Interference may occur in the vicinity of equipment
to 2.7 GHz
marked with the following symbol:
MAXFRAME AUTOSTRUT™ System • Physician User Manual
3,5
]P
V1
12
]P
V2
12
]P
80 MHz to 800 MHz
E1
23
]P
80 MHz to 2.5 GHz
E1
55