4. Emergency Stop Button Information.
The emergency stop button is located on the side of the
control box.
It is a switch that will disable the motors from moving.
The button is there to enable the patient to stop the de-
vice in case of emergency. You should help your patient
to understand what types of situations would be consid-
ered and emergency.
If pressed during active treatment, the motors will stop
moving immediately. The System will beep for 5 seconds
twice and will display the red and orange LED lights. After
that, the red and orange lights will blink for one second
every minute. If pressed while the device is not in active
treatment, nothing will happen until it is time for the next
active treatment to occur. When this time comes, the
device will sense that the emergency stop button has
been pressed and will not start treatment. Instead, it will
display the emergency stop sound and lights described.
■ Note:
Treatment is stopped until system is reactivated by the
physician. The patient should contact the physician as
soon as possible for any situation in which the emergency
stop button was pressed. The patient cannot restart the
device on their own.
If the patient has pressed the emergency stop button,
and the physician has determined that the System should
be started again, a new PDF treatment plan will need
to be created and uploaded to address the residual
deformity.