8.4 Electromagnetic Compatibility (EMC)
Declarations
Guidance and manufacturer's declaration –
electromagnetic emissions
The MAXFRAME AUTOSTRUT System is intended for use in
the electromagnetic environment specified below.
The customer or the user of the MAXFRAME AUTOSTRUT
System should assure that it is used in such an environment.
Emissions test
RF emissions CISPR 11
Harmonic emissions IEC
61000-3-2
Voltage Fluctuations And Flicker
IEC 61000-3-3:2013
Compliance
Electromagnetic environment – guidance
Group1 Class A
The system uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
Class A
The system is suitable for use in all establishments other than
domestic and may be used in domestic establishments and
those directly connected to the public low-voltage power sup-
Complies
ply network that supplies buildings used for domestic pur-
poses, provided the following warning is heeded:
▲ WARNING:
This equipment/system is intended for use by healthcare pro-
fessionals only. This equipment/system may cause radio inter-
ference or may disrupt the operation of nearby equipment.
It may be necessary to take mitigation measures, such as
reorienting or relocating the system or shielding the location.
MAXFRAME AUTOSTRUT™ System • Physician User Manual
53