5 Mounting
Mounting
Flume
42
A false signal suppression with the installed sensor is recom-
mended but not mandatory
The measurement through a ball valve with unrestricted channel
is possible
The deviation can increase in the area of the connecting tube to
the container ± 200 mm
Constructional requirements on the bypass pipe:
Material metal, smooth inner surface
In case of an extremely rough tube inner surface, use an inserted
tube (tube in tube) or a radar sensor with tube antenna
Flanges are welded to the tube according to the orientation of the
polarisation
Gap size with junctions ≤ 1 mm (for example, when using a ball
valve or intermediate flanges with single pipe sections)
Diameter should be constant over the complete length
5.7

Measurement setup - Flow

In general, the following must be observed while mounting the device:
Mounting the sensor on the upstream or inlet side
Installation in the centre of the flume and vertical to the liquid
surface
Distance to the overfall orifice or Venturi flume
Distance to the max. height of the orifice or flume for optimum
accuracy: > 250 mm (9.843 in)
Requirements from approvals for flow measurement, e.g.
MCERTS
Predefined curves:
A flow measurement with these standard curves is very easy to set
up, as no dimensional information of the flume is required.
Palmer-Bowlus flume (Q = k x h
Venturi, trapezoidal weir, rectangular flume (Q = k x h
V-Notch, triangular overfall (Q = k x h
Channel with dimensions according to ISO standard:
When selecting these curves, the dimensions of the flume must be
known and entered via the assistant. As a result, the accuracy of the
flow measurement is higher than with the specified curves.
Rectangular flume (ISO 4359)
Trapezoidal flume (ISO 4359)
U-shaped flume (ISO 4359)
Triangular overfall thin-walled (ISO 1438)
Rectangular flume thin-walled (ISO 1438)
Rectangular weir broad crown (ISO 3846)
The value given takes into account the block distance. At smaller distances,
7)
the measuring accuracy is reduced, see " Technical data".
7)
)
1.86
)
2.5
VEGAPULS 6X • Two-wire 4 ... 20 mA/HART
)
1.5