Exponential Arb (Option E36150Advu) - DATATEC KEYSIGHT E36150 Series User Manual

Autoranging
Table of Contents

Advertisement

Exponential Arb (Option E36150ADVU)

Step 1– Select Arb waveform
Press Arb to access the Arb selection menu. Then select Exponent.
Press Back to return to Arb main menu.
Step 2 – Configure the Exponential Arb
Configure the Exponential Arb accordingly. Refer to the below table for details.
Parameter
Available Key Settings
Arb Type
Voltage or Current
Start Setting (I0 or
minimum to maximum
V0)
Peak Setting (I1 or
minimum to maximum
V1)
Delay (t0)
0 – 3600 s
Time (t1)
0 – 3600 s
Time Constant (tc) 0 – 3600 s
Keysight E36150 Series User's Guide
Sets the output Arb type.
Default is Voltage.
Action required:
Press Arb Type to toggle between Volt and Curr.
Volt
Selects voltage as output Arb type.
Curr
Selects current as output Arb type.
Sets the setting before the waveform.
Sets the end setting of the waveform.
Sets the delay time after the trigger is received but before the waveform starts.
Sets the time for the amplitude to go from the start setting to the end setting.
Sets the time constant of the curve.
Description
99

Advertisement

Table of Contents
loading

This manual is also suitable for:

E36154aE36155a

Table of Contents