DATATEC KEYSIGHT E36150 Series User Manual page 94

Autoranging
Table of Contents

Advertisement

Parameter
Available Key Settings
Voltage after Arb
DC or LAST
Current after Arb
Trigger Source
Key, IO, or Rmt
Repeat Count
1 to 16,777,216
Continuous
On or Off
Edit Points
-
Step 3 – Run the Arb
When the arbitrary waveform runs, front panel voltage and current controls as well as any remote voltage and
current commands are ignored until after the Arb completes.
94
Specifies what happens to the parameter settings when the Arb completes.
Default is DC.
Action required:
Press After Arb to toggle between DC and LAST.
DC (Return to DC
The parameter setting returns to the DC value that was in effect prior
Value)
to the Arb.
LAST (Last List
The parameter setting remains at the last Arb value.
Value)
Sets the trigger source for the system.
Default is Rmt.
Action required:
Press Properties > Trig Src>to toggle between Key, IO, and Rmt.
Key (Arb Run/Stop Key)
IO (DIO Trigger In)
Rmt (Remote Command)
Sets the Arb repeat count. This sets the number of times that an Arb is executed before it
completes.
Default is 1.
Enables check box to repeat the Arb continuously.
Default is On.
Action required:
Press Properties > Continuous to toggle between On and Off.
On
Enables the check box.
Off
Disables the check box.
Create a user-defined Arb from the present Arb property values. This lets you edit specific
points in the standard arbitrary waveform.
Description
Selects the Arb Run Stop key or [List Run/Stop] as
a trigger source.
Selects any configured digital IO (with Trigger
Input function) as a trigger source
Selects a remote interface command as a trigger
source.
Keysight E36150 Series User's Guide

Advertisement

Table of Contents
loading