Specifying 2-Wire Or 4-Wire Sense - DATATEC KEYSIGHT E36150 Series User Manual

Autoranging
Table of Contents

Advertisement

Specifying 2-Wire or 4-Wire Sense

When 4w mode is selected, ensure the load is connected to the rear panel output for accurate reading.
DO NOT USE both front and rear panel output concurrently. This may lead to inaccurate output reading.
You have the option to use the remote sense terminals on the rear panel to measure the voltage at the device under
test (DUT).
1. Press Source Settings.
2. Press Sense 2w or Sense 4w to select 2-wire sense or 4-wire sense.
3. Press Back to exit and return to the meter-view display.
The 2w mode closes a relay within the power supply to short the output and sense connectors. This means that only
two wires are used, and 4-wire sense is disabled. The 4w mode opens the relay in order to separate the output and
4-wire sense inputs. By default, 2w is selected.
From the remote interface:
To set the remote sense relay to 4-wire sense:
VOLT:SENS EXT