Pc Control - BRUEL & KJAER 5998 Service Manual

Reciprocity calibration apparatus
Table of Contents

Advertisement

Acceptor
handshake
Listener
Remote Local
Device clear
All other functions: no capability
Remote/Local
You can control the multiplexer either locally or remotely from a PC. To switch
between the two states:
Local → Remote: send a command
Remote → Local: press <Local> on the front panel of Type 5998
The current state of the reciprocity calibration apparatus is indicated by the Re-
mote LED on the front panel.
Commands
Type 5998 is remotely controlled by sending a string which sets all required param-
eters. If a parameter is not included, the value is not changed. The commands are
listed in
Note: A letter P must be sent before each parameter
Example
P0OP2GP0RP1F
Output A, 20 dB gain, reference signal off, 20 Hz high-pass filter.
5.2.4

PC Control

Control of the system using the Brüel & Kjær measurement program BZ 5226 is
described in the associated on-line help. See also
BI 5044 –11
Function
AH1 – complete
compatibility
L2 – basic listener
RL2 – no local lockout
DC2 – omit selective
device clear
Table 5.1 IEEE– 488 functions implemented
Table
5.2.
Reciprocity Calibration Apparatus Type 5998
Subset
Clause
Service Manual
Chapter 5 — Theory of Operation
Functional Description
IEEE/
IEC
ANSI
Section
7
2.4
9
2.6
11
2.8
13
2.10
section
4.2.