Maximum Voltage And Distortion, Channel A - BRUEL & KJAER 5998 Service Manual

Reciprocity calibration apparatus
Table of Contents

Advertisement

3.3.3

Maximum Voltage and Distortion, Channel A

1. Set the front panel controls on Type 5998 as follows:
2. Apply about 2.2 V at 1 kHz to the Channel A Receiver input using cable
3. Adjust the generator output until the Overload indicator just lights.
4. Use the oscilloscope to check that the signal on the Output BNC socket is not
BI 5044 –11
Control
Gain
Output
HP Filter
Ref.
Polarization Voltage
AO 0479.
clipped.
Reciprocity Calibration Apparatus Type 5998
Setting
10
A
20
Off
0
Service Manual
Chapter 3 — Checking Procedure
Checking Procedure