Chapter 3 — Checking Procedure
Checking Procedure

3.3.10 Reference Signal

1. Set the front panel controls on Type 5998 as follows:
2. Apply 3.5 V at 1 kHz to the Generator Input BNC socket using cable AO 0479.
3. Use the Measuring Amplifier to check there is about 35 mV on the Output BNC
4. Use the Measuring Amplifier to check there is the same 35 mV on:
5. Check that the Ref. Adj. potentiometer can change the signal about ± 10 dB.
3 − 12
Control
Gain
Output
HP Filter
Ref.
Polarization Voltage
socket (– 40 dB gain) when the Ref. Adj. potentiometer is in its mid-position.
pin 1 of the Channel A Receiver BNC socket and
pin 1 of the Channel B Transmitter BNC socket and
on the Ref. Signal Output BNC socket on the rear panel.
Reciprocity Calibration Apparatus Type 5998
Setting
0
Ref.
20
On
0
Service Manual
Brüel & Kjær