Assembly Of The Display Stand; Assembly Of Flange Mounted Transducers To Transducer Plate - norbar TWL 1500 Operator's Handbook Manual

Torque wrench loader
Table of Contents

Advertisement

C.

ASSEMBLY OF THE DISPLAY STAND

1)
Screw the display stand to the display stand base using 2 off M5x16mm long
screws.
2)
Insert the display stand base into the display support arm and tighten using the
thumb nut from the display stand bolt kit.
3)
Attach the support arm to the transducer plate and fasten using the M10x50mm
long screw and thumb nut from the display stand bolt kit.
DISPLAY
STAND
M10
SCREW
D.
ASSEMBLY OF FLANGE MOUNTED TRANSDUCERS TO TRANSDUCER
PLATE.
1)
Place the spacers for the smaller transducers in the position shown below.
2)
Attach the transducers to the transducer plate in the positions shown below using
the bolts supplied in the flange mounted transducer bolt kit. Tighten the
M5x75 bolts to 5Nm, the M8x60 bolts to 10Nm and the M12x30 bolts to 15Nm.
3)
Ensure that the cables are extending out to the display stand side of the unit.
4)
Connect the transducer lead to the Torque Tool Tester display instrument and
mount it on the display stand. Connect the power supply to the Torque Tool
Tester.
SPACERS
THUMB
NUTS
FMT1500
FMT150
FMT400
PAGE 5 OF 8
Issue 2
JANUARY 2005
M5
SCREWS
DISPLAY
STAND
BASE
DISPLAY
SUPPORT
ARM
FMT10
FMT25

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents