ZOLL Infusion Dynamics Power Infuser M100B-3A Operating Instructions Manual page 18

Table of Contents

Advertisement

Routine Maintenance
keep clean
After each use, clean dirt
and salt solutions from
the top of the pump
by flushing with
water and drying
thoroughly. Salt solu-
tions can defeat the
pump's air detection system
and salt crystals can damage the
pump's seals. Clean foreign matter from
under the pump rocker to prevent jamming.
Do not allow water or IV fluids
to enter the battery compart-
ment.
Do not manipulate the
pump rocker with your fin-
ger—it can damage the
mechanism.
Annual performance testing is recommended
for organizations that wish to test the pump
on a periodic basis to ensure that it is calibrat-
ed and functioning to specifications. Contact
Infusion Dynamics for a copy of the End-
User Performance Test Procedure (0080-
0226).
17

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents