catheter. This feature can add flexibility
in the way you use the pump. For exam-
ple, use gravity to generate a low keep-
vein-open rate or continue fluid delivery
when the pump is disconnected. However,
be aware that free flow can cause the
actual flow rate to be higher than the set
rate, particularly at the low settings.
In the event of a pump failure or unin-
tended free flow, the maximum volume of
fluid that can be infused will be deter-
mined by the volume remaining in the IV
bag (typically 1 liter maximum).
Air Elimination (crystalloid/colloid car-
tridge only). Filters in the crystalloid/colloid
pump cartridge will vent air from the IV line
once the cartridge has been primed.
Do not use blood or blood products with
the Crystalloid/Colloid Cartridge—the 1.2
micron filter will clog.
These membranes are not included in the
Blood Cartridge. When using the Blood
Cartridge, PURGE ALL AIR when prim-
ing and before starting the infusion.