Electromagnetic Compatibility (Emc) - Thermo Scientific TDE Series Manual

Ultra-low temperature freezers
Hide thumbs Also See for TDE Series:
Table of Contents

Advertisement

ELECTROMAGNETIC COMPATIBILITY (EMC)

Guidance and Manufacturer's Declaration—Electromagnetic Emissions
Emission Test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class A
CISPR 11
Guidance and Manufacturer's Declaration—Electromagnetic Immunity
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions, and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
4 |
Safety Considerations
Electromagnetic Environment Guidance
The EMISSIONS characteristics of the freezer make it suitable for use in commer-
cial, industrial or institutional occupancies (CISPR 11 class A). If it is used in a resi-
dential environment (for which CISPR 11 class B is normally required), this
equipment might not offer adequate protection to radio-frequency communication
services. The user might need to take mitigation measures, such as relocating or
reorienting the equipment.
IEC 60601-1-2 Test
Level
±8 kV contact
±15 kV air
±1 kV for input/output
lines ±2 kV for AC
powerlines
±1 kV differential mode
±2 kV
common mode
0% UT (100% dip in
UT) for 0.5 cycle at 0°,
45°, 90°, 135°, 180°,
225°, 270° and 315°
.0% UT (100% dip in
UT) for 1 cycle at 0°
70% UT (30% dip in
UT) for 25/30 cycles
at 0°
0% UT (100% dip in
UT) for 250/300 cycles
30 A/m
Compliance Level
±8 kV contact
±15 kV air
±1 kV for input/output
lines ±2 kV for AC
powerlines
±1 kV differential mode
±2 kV
common mode
0% UT (100% dip in
UT) for 0.5 cycle at 0°,
45°, 90°, 135°, 180°,
225°, 270° and 315°.
0% UT (100% dip in
UT) for 1 cycle at 0°
70% UT (30% dip in
UT) for 25/30 cycles
at 0°
0% UT (100% dip in
UT) for 250/300 cycles
30 A/m
Electromagnetic Environment Guidance
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%
Mains power quality should be that of a
typical commercial or industrial environ-
ment.
Mains power quality should be that of a
typical commercial or industrial environ-
ment
Mains power quality should be that of a
typical commercial or industrial environ-
ment
Mains power quality should be that of a
typical commercial or industrial environ-
ment.
Thermo Fisher Scientific ULT Freezers

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents