Thermo Scientific TDE Series Manual page 42

Ultra-low temperature freezers
Hide thumbs Also See for TDE Series:
Table of Contents

Advertisement

S.No
Parameter
Build
22
Number
Cabinet
23
Calibration
Offset
24
Line Voltage
Thermo Fisher Scientific ULT Freezers
Function
Address in
Size
Code
hexadecimal
0x03
524
2
0x03
534
2
0x03
4F8
2
Relay
Modbus
Enclosure
Data
Response
3a 31 39
3A 31 39
30 33 30
30 33 30
35 32 34
34 30 30
0x00
30 30 30
000D
30 30 30
01
32 42 39
44 30 31
0d 0a
43 45
0D 0A
3a 31 39
3A 31 39
30 33 30
30 33 30
35 33 34
34 30 30
30 30 30
0xFF
30 30 30
FFFC
32 41 39
30 30 30
E0
0d 0a
44 43
0D 0A
3a 31 39
3A 31 39
30 33 30
30 33 30
34 30 30
0x00
34 46 38
0000
30 30 30
30 30 30
E4
32 45 36
30 45 34
0d 0a
46 34
0D 0A
Appendix B: Modbus ASCII Parameter Table |
Data
Remarks
Type
XX/XX Minor (numbers after
decimal point)/ Major
(numbers before decimal
unit
point)
Ex: 1301 translates to Build
number is 1.13
Convert the data value into
signed 2's complement and
divide with 10,which gives
the setpoint.
int
Ex: Signed 2's complement
of the 0xFFFFFCE0 is equal
to –800.
–800/10=–80.So the
setpoint is –80C.
Convert hex to decimal
gives the voltage.
Ex. Here 0xE4 is equal to
unit
228 V
37