6-3 Percent Mode (%)
The percent mode displays the weighting value in percentage compared to a 100%
reference mass and is used for target weighing or checking the sample variance.
Selecting The Percent Mode
1. Press the MODE key to select the unit % (Percent mode).
Storing The 100% Reference Mass
2. Press the SAMPLE key to enter the 100% reference
mass storing mode.
Even in the storing mode, pressing the
switch to the next mode.
3. Place a container on the weighing pan, if necessary. Press the RE-ZERO
key to cancel the weight (tare). The balance displays 100 0 % .
4. Place the sample to be set as the 100% reference mass on the pan or in
the container.
5. Press the PRINT key to store the reference mass.
The balance displays 100. 0 0 %. (The decimal point
position depends on the reference value. The reference
mass stored, even if the AC adapter is removed, is
maintained in non-volatile memory.)
Note
□ If the balance judges that the mass of the sample is too light to be
used as a reference, it displays
□ The displayed percentage is based on the 100% reference mass.
Model
Minimum display 0.01g
Minimum display 0.1g
□ Registered values are stored even when the power is turned off.
6. Remove the sample
Reading the percentage
7. Please a sample to be compared to the reference mass on the pan.
The displayed percentage is based on the 100% reference mass.
MODE key will
.
lo
100% mass
1.00g ~ 9.99g
1%
10.00g ~ 99.99g
0.1%
100.00g ~
0.01%
10.0g ~ 99.9g
1%
100.0g ~ 999.9g
0.1%
1000.0g ~
0.01%
28
Decimal point position
Weighing pan
Container
Sample
(100%)
place
Display % of weighing object