6 Detailed specifications
Output sensitivity ( FS )
Maximum capacity ( Emax )
Max.number of load cell
intervals
Ratio of min. LC verification
interval
Combined Error
Minimum dead load
Safe overload
Ultimate overload
Zero balance
Excitation, recommended
voltage
Input resistance
Output resistance
Insulation resistance
Compensated temperature
Operating temperature
7 Wiring
Shielded, 6 cable. Shield not connected to element.
Cable diameter: Ø6mm.
Standard cable length for 300kg – 5000kg: 5m
Standard cable length for 7500kg – 10000kg: 10m
6-wire diagram
4
CT Q1-BA-e-1810
mV/V
300, 500, 750, 1000, 1500, 2000, 3000,
kg
nLC
Y = Emax
/ vmin
%FS
Of Emax
of Emax
of Emax
of FS
V
Ω
Ω
MΩ
°C
°C
Input(+)
Green
Sense(+)
Blue
Output(-)
White
Sense(-)
Yellow
Black
Input(-)
Output(+)
Red
2.0 ± 0.001
5000, 7500, 10000
3000
10000
≤± 0.017
0%
150 %
200 %
< ± 2.0 %
10 ~ 15
400 ± 20
350 ± 3
≥5000 ( at 50VDC )
-10 ~+40
-35 ~ +65