Patient-to-Ground Leakage for FECG
Revision C
Maintenance: Preventative Maintenance Inspection
1.
Connect an FECG test body to the monitor's front panel
2.
Configure a leakage tester to perform a patient leakage test.
3.
Turn ON the monitor using the monitor's
4.
Verify the following:
<50 µA
(132–253 VAC, or at the line voltage at your facility +10%)
5.
Record the patient-to-ground leakage for the following conditions:
Table 9-10. Patient-to-Ground Leakage Conditions for FECG
Neutral
__________ Closed
__________ Closed
__________ Closed
__________ Closed
__________ Open
__________ Open
170 Series Monitor
2000947-004
button.
On/Standby
Ground
__________ Closed
__________ Normal
__________ Open
__________ Normal
__________ Open
__________ Reversed
__________ Closed
__________ Reversed
__________ Closed
__________ Normal
__________ Closed
__________ Reversed
input.
FECG
Power (Polarity)
9-17