WACON AquaInform Sycon 2500 Operating Manual page 48

On-line water hardness monitor
Table of Contents

Advertisement

P
48
AGE
D
10
IAGNOSIS
10. X
START
D
11
IAGNOSIS
11. X
START
DANGER
D
12
IAGNOSIS
12. X
START
FLUSH
NOTE
– D
S
C
YSTEM
HECK
IAGNOSTIC
C
A
HECK
GITATOR
th
After 10
pressing the START button, the
function of the agitator is reviewed.
The red + blue LED will flash and the
agitator will switched on
C
D
I
IN
HECK
IGITAL
NPUT
th
After 11
pressing the START button, the
function of digital input contact is reviewed.
terminal 13 + 14
If the digital input contact is bridged,
the left red + yellow LED light on.
If the digital input contact is opened,
the left green + yellow LED light on.
It is recommended to close the lid of the control box before running
through diagnosis step 12 and 13
C
Z
T
HECK
ERO
RANSMISSION
th
After 12
pressing the START button, the
function of zero transmission measurement
in the optical path of the mixing chamber is
reviewed.
the first three LEDs from the left green
+ yellow + blue light up
For the examination of the test section,
the measuring chamber should be
filled
with
clean
measuring chamber by repeatedly
pressing the FLUSH button
This test step is required to perform
the zero value of the sample for the
following examination of the colour
detection
→ step 13
Note that for a correct measurement
switches S6, S7 and S8 are in the right
position
according
parameters (→ page 32).
F
UNKTIONS
water.