WACON AquaInform Sycon 2500 Operating Manual page 17

On-line water hardness monitor
Table of Contents

Advertisement

C
2
HAPTER
L
A
IMIT
NALYSER
1mg/L
0.05 °dH
S
PECIFICATION
2.6 Functions
The SYCON 2500 system has the following features:
[1]
Monitors hardness leakage automatically in accordance to limit value
indicator used.
The monitoring process is fully automated, saving a significant amount of
work by eliminating the need for complicated manual procedures.
[2]
Requires no periodic calibration
The system needs no cumbersome periodic calibrations.
[3]
Interval for each monitoring may be set in 4 programmable intervals: 5 / 10
/ 20 / 30 minutes
[4]
Reliable detection of hardness leakages by use of limit value Indicators
2mg/L
4mg/L
0.10 °dH
0.20 °dH
[5]
Evaluates hardness leakage at higher accuracy
When a BAD-condition is detected, monitoring may be repeated with a
delaytime of 4 minutes
[6]
LED status display is independent on language areas
[7]
Offers an alarm function
When it is evaluated that there is a hardness leakage, an alarm is issued
by switching a potential free relay. This remote alarm output may be used
to send an alarm remotely or sounds a buzzer for example.
[8]
Offers a diagnostic-program. If a problem occurs in the system an alarm is
issued by switching a potential free relay. Technicians may run step by
step through the diagnostic program to check functions or find faulty parts
(→ page 44).
in the device
[9]
Requires minimal maintenance
Depending on measurement intervals and frequencies respectively the
measuring chamber has to be cleaned up and indicator hose pipes have to
be changed app. twice a year
[10]
Minimal indicator consumption
The indicator bottle may be replaced easily. The 500 ml bottle needs no
replacement for approximately three to four months in typical applications.
(Note that more frequent replacement may be necessary, depending on
the application.)
[11]
Compact in design, easy to install
The main unit is installed easily on a wall.
Installation is a simple process
Wacon provides the most compact designed devices of this type.
6mg/L
10mg/L
20mg/L
0.30 °dH
0.50 °dH
1 °dH
(→ page 38).
(→ page 19)
.
P
17
AGE
40mg/L
100mg/L
200mg/L
2 °dH
5 °dH
10 °dH

Advertisement

Table of Contents
loading

This manual is also suitable for:

30-010120

Table of Contents