3.3 Installation in 4 steps
The analyser can be mounted with or without protective housing. The
manufacturer offers a standard housing; the mounting and dimensions are
described in this guide:
Custom-built or tailor made enclosures for the devices of the series aqua inform
LimitAnalyser / aqua inform TrendAnalyser / SYCON 2500 / SYCON 2800 and
SYCON 3000 as well as devices with custom names and labels that are
technically based on these ranges are not described in detail in this manual.
►
step I a
mounting without enclosure
Using 4 screws (max. 6 mm) to mount the unit on a wall or suitable support
structure. Hole spacing see page 24
►
step I b
Using the included 4 brackets to mount the unit (→ page 24). The tabs can be
rotated by 45 degrees or 90 degrees outside. Alternatively, the unit can be
mounted without the brackets from the rear on a plate (M6). For hole spacing see
page 24
In both variants, avoid direct sunlight and strong artificial light sources
→ disturbance of the optical path in the measuring chamber
→ electromagnetic interferences by artificial light source
Do not install under dripping pipes.
►
step II
eatablish sample and drain pipe line
Use flexible hose pipe 6x4. Between water treatment plant and analyser a
manual shutoff valve and particle filter (if necessary) should be installed.
Drainage should be lead via short connection into an open funnel → runoff to
atmospheric pressure.
Make sure that you connect the inlet and outlet properly
→ Figure 2.1 in page 13
►
step III electrical connections
refer to information in figure 3.1 (terminal connection (→ page 21) and figure 3.2
(connection instructions (→ page 22).
Work on electrical equipment of the plant / machinery must be performed by a
qualified electrician! Observe all applicable electrical installation work rules
→ applicable supply voltage 85 ... 264 V AC 47 ... 63Hz