WACON AquaInform Sycon 2500 Operating Manual page 41

On-line water hardness monitor
Table of Contents

Advertisement

C
5
M
HAPTER
S
2500
YCON
M
R
AINTENANCE
S
3
TEP
1. remove empty indicator bottle
2. replace full indicator bottle
NOT
E
3. de-aerate the indicator hose pipe
FLUSH
INDICATOR
4. confirm bottle change and reset counter
+
AINTAINANCE
I
R
EFILL
NDICATOR AND
ESET THE
→ open bottle cap by turning
→ remove the suction pipe carefully
→ If indicator fluid leaks, wipe up with a paper
towel
→ dispose of empty bottle
→ remove the cap
→ insert the suction pipe carefully
→ screw bottle cap by turning
When you are installing pay attention to a
correct position of the screw and the adapter.
→ fill measuring chamber with water
→ press FLUSH button several times
→ dose indicator until the hoses and the
cartridge of peristaltic pump are completely
filled with indicator and no air bubbles are
visible
→ press INDICATOR button several times
→ press
INDICATOR
simultaneously until the right red and the blue
LED light up briefly
S
C
TORAGE
OUNTER
and
RESET
button
P
41
AGE
blue
red