4.3 Setting the program switches
The Limit Analyser is programmed using small dual-in-line-switch S1-S10 and
adjusted to specific operating requirements.
switch off the unit and open the lid of he controller
→ WARNING supply voltage 85 ... 264 V AC 47 ... 63Hz
WARNING
factory settings
flushing time
analysis interval
retry cycle
relay function
parameter
operating mode
Work on electrical equipment of the plant / machinery must be performed
by a qualified electrician! Observe all applicable electrical installation
work rules.
ELECTRICITY
→ WARNING supply voltage 85 ... 264 V AC 47 ... 63Hz
For the operation of the rotary potentiometer and for setting the slide
switch you need a small insulated screwdriver
NOTE
S
PERATING THE
YSTEM
→ The DIP-switches (dual in-line package) are located between
the LED lights and the push buttons. Next to the switches
you see the potentiometer for setting the flushing time.
S P Ü L Z E IT
F
T
LUSHING
IME
ca. 2 minutes
10 minutes
yes
permanent
without analysis stop
total hardness
analysis mode
Please use only good, proven tools, you will help avoid damage to
sensitive components
O N
1
2
3
4
5
6
7
D
S
IP
WITCHES
Poti app.
S1 OFF
S3 ON
contact
S4 ON
S6 OFF
S9 OFF
P
29
AGE
8
9
1 0
10:00
S2 ON
S5 OFF
S7 OFF
S8 OFF
S10 OFF