WACON AquaInform Sycon 2500 Operating Manual page 33

On-line water hardness monitor
Table of Contents

Advertisement

C
4
O
HAPTER
L
A
IMIT
NALYSER
4.4
Make sure that the analyser was installed in accordance with chapter 3
(→ page 19) and the program switches were set properly according to the
application like described under Section 4.3 (→ page 29).
HINWEIS
step 1
switch on the unit
→ switch power button „on"
→ green LED
step 2
reset indicator storage volume
→ Press the INDICATOR and RESET
+
step 3
flood measuring chamber with water
→ Press the FLUSH button until the
step 4
de-aerate dosing pump
→ Press the INDICATOR button until
step 5
start analysis
→ press START button to run first
→ an analysis starts with the flushing of
S
PERATING THE
YSTEM
Start up in 5 steps
"analysis result" flashes.
button at the same time
WARNING !!I
run reset function only when replace
empty bottle by a full one
(→ page 41)
measuring chamber is completely filled
with water and free of bubbles.
indicator is delivered continuously into
the measuring chamber. During this
process, the agitator rotates.
analysis
the measuring chamber
P
AGE
green
INDICATO R
R ESET
FLUSH
INDICATO R
ST AR T
33