WACON AquaInform Sycon 2500 Operating Manual page 39

On-line water hardness monitor
Table of Contents

Advertisement

C
5
M
HAPTER
S
2500
YCON
► Maintenance in 3 steps
step 1 - step 2A without 2B - step 3
routine maintenance in intervals of app. 4-6 month
requirements
time
material
tools
prior to
maintenance
MAINTENANCE
1
STEP
After app. 6 month, the pump cartridge with
hose pipe should be replaced because the the
terminal clutch wears out and tube material
becomes brittle.
order no.
33-090038
1. Press locking clips together with thumb and
2. open the bayonet connectors of the LUER-
3. If indicator fluid leaks, wipe up with a paper
3. replace new peristaltic pump cartridge in
4. de-aerate pump
INDICATOR
AINTAINANCE
app. 30 minutes
indicator → page 12
maintenance set no.1
cleaning set
paper towels
screw driver
cup or small bucket
cleaning set
S
WITCH OFF THE UNIT
►switch power off
C
C
P
HANGE
ARTRIDGE OF
description
cartridge of peristaltic pump
forefinger and pull cartridge to the right off
the motor shaft
fittings
towel.
reverse order
→ press button INDICATOR
depending on limit
Art.nr. 33-090 038