Empatica EmbracePlus Instructions For Use Manual page 128

Hide thumbs Also See for EmbracePlus:
Table of Contents

Advertisement

1. This device may not cause harmful interference, and
2. This device must accept any interference received, including
interference that may cause undesired operation (FCC Title 47,
Subpart A, Part 15.19(3)).
Changes or modifications not expressly approved by the party
responsible for compliance could void the user's authority to operate
the equipment (FCC Title 47, Subpart A, Part 15.21)
GUIDANCE AND MANUFACTURER'S DECLARATION -
ELECTROMAGNETIC IMMUNITY
The EmbracePlus is intended for use in the electromagnetic
environment specified below. The user of the EmbracePlus should
assure that it is used in such an environment.
Immunity Test
IEC
test level
Electrostatic
8
Discharge
contact
(ESD)
EN
2/4/8/15
61000-4-2
kV Air
Radiated
10 V/m
electromagn
80 MHz to
etic field
2.7 GHz
IEC
61000-4-3
60601
Compliance
Level
kV
IEC
60601-1-2
Test level
IEC
60601-1-2
Test level
Electromagnetic
guidance
Floors should be wood,
concrete, or ceramic tile. If
floors are covered with
synthetic
material,
relative humidity should
be at least 30 %
Portable and mobile RF
communications
equipment should be used
no closer to any part of
EmbracePlus,
environment-
the
including
127