Power supply
Battery
power supply from 1 or 2 exchangeable NiCd
batteries
rated voltage 12 V
rated capacity 2.0 Ah
battery charging in device
max. charge time of a depleted battery 4 hours
operating time with one fully charged battery
approx. 60 shocks of 360 Joules each (into 50
Ohms) or 1.4 hours of monitoring (1 h if pace-
maker, etCO
and SpO
2
used)
Line Power (optional)
100 V to 240 V ± 10 % (47 Hz to 63 Hz)
power consumption at 230 V:
− during defibrillator charging 240 VA
− during monitor operation 40 VA
− during battery charging 25 VA
max. charge time of a depleted battery 4 hours
intrahospital use (IPX 1)
Operation in Vehicle Mounting Unit, Power
from 12 V DC Ambulance Supply System:
11.5 to 18 V
power consumption
− during defibrillator charging 130 W
− during monitor operation 27 W
− during battery charging 14 W
Operational readiness
4 seconds, incl. automatic self-test
Operating position
any, horizontal with AC power adapter
227 490 02-C
Technical Specifications
measuring system are
2
Marquette Responder® 3000
Type of Protection
protected against splashing water; Marquette
Responder: IPX 4 (batteries inserted, cables
connected)
AC power adapter: IPX 1
Environment
Operation
operation under the following conditions consid-
ered to be normal:
temperature between 0 and +45 °C
relative humidity between 5 % and 95 %, no
condensation
atmospheric pressure between 700 and
1060 hPa
vibration acc. to MIL 810E Cat.10 and
prEN 1789
Transport and storage
temperature between -25 and +70 °C
relative humidity between 10 % and 95 %, no
condensation
atmospheric pressure between 500 and
1060 hPa (with etCO
1060 hPa)
Dimensions
width 307 mm
height 202 mm
depth
412 mm (without AC power
adapter)
476 mm (with AC power adapter)
Weight
with one battery approx. 8 kg (w/o. AC power
adapter)
adapter cable: 700 and
2