GE Marquette Responder 3000 Operator's Manual page 56

Table of Contents

Advertisement

a
d
Figure 10-3. Attaching the airway adapter
89
I
160 / 40
ECG
bpm
II
37
---
20
etCO2
mmHg
III
15.07.1998
09:05:00
QRSPulse
etCO2
ECG
Tone
SpO2
OFF
Figure 10-4. etCO
reading
2
High
Alarm
Low
etCO2
etCO2
Figure 10-5. Alarm limits menu
Note
etCO
alarms are similar to HR alarms. Please
2
refer to chapter 7 "Displaying and Monitoring the
ECG".
56
Capnometry (etCO
c
b
semiautom.
Next
Menu
Alarm
Previous
Menu
Marquette Responder® 3000
)
2
Measuring and Monitoring etCO
Obtain a new airway adapter and check that the
windows are intact and clean.
Attach the airway adapter to the sensor (the
triangles on the airway adapter and on the
sensor must be in alignment - Figure 10-3).
Connect the airway adapter to the respirator
system (connector c towards the patient, con-
nector d towards the respirator - Figure 10-3).
Ensure that the sensor is on top and that the
windows of the airway adapter are in a vertical
position. Although the "No-fog-membrane"
prevents moisture condensation on the window,
liquid secretions could soil the window.
Connect the sensor to the defibrillator (yellow
connector, right).
The CO
measurement begins as soon as the sensor
2
is connected. During the short self-test, the
parameter window is blue. As soon as the sensor
supplies a valid reading, the color changes to
green. Check that the indicator on the CO
cable lights green; if a problem is detected, the
indicator flashes red (apnea or system failure).
At the factory, only a low etCO
is adjusted, no high limit. When the parameter
reading drops below this value, the defibrillator
sounds an alarm and the window changes from
green to red.
Adjusting Alarm Limits
The alarm limits can be modified permanently
from the setup menu or temporarily as described
below:
Press F2
SpO2 etCO2
Press F1
etCO2 Alarm Limits
limits menu (Figure 10-5).
Change the high limit with F1, F2 and the low
limit with F3, F4.
To deactivate the etCO
the sensor connector and press F2
2
adapter
2
limit of 20 mmHg
2
.
to display the alarm
measurement, remove
2
SpO2 etCO2
227 490 02-C
.