Procedure:
1. Load the cal kit file into your computer. The cal kit file is located on the
2. Select Cal > Cal Sets & Cal Kits > Cal Kit...
3. Select Import and open the WR12_VDI calibration kit file on the USB
4. Select Preset
5. Confirm system is in mm-wave mode frequency and setup the number
6. Select Cal > Other Cals > Smart Cal
7. Select the box for "Calibrate Power" (see Figure 28 below) if you have a
8. Due to the RF and LO multiplication, you may need to calibrate the
Figure 29
Smart Cal GUI
9. Refer to the Help menu to complete the Smart Cal.
10. Once the Smart Cal is complete, save the calibration if desired.
For 1-Port systems, Smart Cal GUI will show two ports (see
on page
calibration to not complete.
N5252-90002 User's Guide
USB drive that came with the cal kit. Plug the USB drive into the USB
port on your PC.
(.ckt or .xkt).
of points in sweep (default is 201).
power meter and E-Band sensor or are using the power data from the
extender USB.
VNA source to get the best leveling and power control.
40). Only select Port 1. Selecting both ports will cause the
Figure 30