N5253E2 Configuration

1. Install the RF/LO and IF cables from Ports 1 and 2 of the P5004A and P5008A to VDI Extenders
#1 and #2. Install RF/LO cable from LO Out of P5004A to LO-I of VDI Extender #2. See
and
Figure
16. Use 11904C adapters on the P5008A.
2. If not already installed, install a 50 Ω load to LO In on P5008A and LO-O on VDI Extender #1.
See
Figure
16.
3. Connect VDI Extender #1 (LO-I) to VDI Extender #2 (LO-O) with an RF/LO cable. See
and
Figure
16.
4. Connect the power cords to VDI Extenders. See
Cable Connections for P5004Aand P5008A to VNAX Extenders
Table 12
Figure
Cable
1
Reference
Type
Blue
RF/LO
Blue
RF/LO
Red
IF
Red
IF
Green
RF/LO
Green
RF/LO
Black
Control
Black
SMA
Figure 16
N5253E2 Configuration
1. RF/LO cables are thicker then the IF cables and are marked RF/LO or IF at the end of the cables.
2. Connections to the P5008A require 11904C adapters.
N5252-90002 User's Guide
Figure
P5004A
P5008A
VNA
2
VNA
Port 1
Port 1
Port 2
Port 2
LO Out
CTRL S
to
CTRL M
LO In
to
LO Out
16.
VDI
Extender #1
to
--
to
RF/R
to
--
to
M
to
--
to
LO-I
Table 12
Table 12
VDI
Extender #2
RF/R
--
M
--
LO-I
LO-O