Summary of Contents for caron CRSY102

  • Page 1 CRSY102 INST-PM-CRSY102...
  • Page 2: Revision Log

    Revision Log Revisions Description Letter Date 01-03-2017 Initial release. New cover. 04-11-2018...
  • Page 3 Parts Required: (1) Carbon Filter, (1) DI Cartridge, (1) Antimicrobial Copper Rod, (1) Atomizer, (1) Silicone Grease Tools Required: (2) 5/8″ Wrenches, Bucket Parts included in Preventative Maintenance Kit for Model CRSY102 are shown below. Carbon Filter DI Cartridge Silicone Grease...
  • Page 4 3. Remove left and right access panels. Remove panels by pulling down on latch tabs, rotating panel slightly outward and lifting as shown below. Access Panel Latch Tabs 4. Remove fastening screw and slide filter wrench off DI cartridge housing as shown below. Filter Wrench Fastening Screw INST-PM-CRSY102 Rev B 04/11/2018...
  • Page 5 Fastening Screw 5. Unscrew DI cartridge housing on left side of recirculator using filter wrench as shown below. Note: Drain water from DI cartridge housing into bucket. DI Cartridge Housing INST-PM-CRSY102 Rev B 04/11/2018...
  • Page 6 Filter Wrench Do not discard O-ring inside of DI cartridge housing. 6. Remove and discard old DI cartridge. INST-PM-CRSY102 Rev B 04/11/2018...
  • Page 7 Note: Do not use more than half of silicone grease. Ensure tapered end of DI cartridge is located next to O-ring groove. Ensure O-ring is seated inside groove in DI cartridge housing. Silicone Grease Tapered End of DI Cartridge O-ring Groove INST-PM-CRSY102 Rev B 04/11/2018...
  • Page 8 1. Unscrew carbon filter housing on right side of recirculator using filter wrench as shown below. Note: Drain water from carbon filter housing into bucket. Carbon Filter Housing Do not discard O-ring inside of carbon filter housing. 2. Remove and discard old carbon filter. INST-PM-CRSY102 Rev B 04/11/2018...
  • Page 9 1/8 turn further as shown below. Carbon Filter Housing 5. Slide filter wrench onto DI cartridge housing and fasten screw. REPLACING ANTIMICROBIAL COPPER ROD 1. Unscrew reservoir cap located on top of recirculator as shown below. Reservoir Cap INST-PM-CRSY102 Rev B 04/11/2018...
  • Page 10 3. Discard old antimicrobial copper rod and spring clip. 4. Slightly squeeze new spring clip together. Then insert long white end of new antimicrobial copper rod into new spring clip as shown below. Antimicrobial Copper Rod INST-PM-CRSY102 Rev B 04/11/2018...
  • Page 11 6. Squeeze spring clip while inserting antimicrobial copper rod as shown below. Antimicrobial Copper Rod 7. Place new antimicrobial copper rod inside reservoir and replace cap. REPLACING ATOMIZER 1. Locate lower reservoir on left side of recirculator as shown below. Lower Reservoir INST-PM-CRSY102 Rev B 04/11/2018...
  • Page 12 2. Unsnap (3) lid tabs on lower reservoir and lift to locate atomizer as shown below. Lid Tabs Atomizer INST-PM-CRSY102 Rev B 04/11/2018...
  • Page 13 4. Install new atomizer by turning nozzle clockwise using 5/8″ wrenches. 5. Reinstall lid and snap (3) lid tabs in place as shown below. Note: Do not snap tab located under filter wrench as shown below. Filter Wrench Snapped Lid Tabs INST-PM-CRSY102 Rev B 04/11/2018...
  • Page 14 Chamber drain/recirculator inlet is gravity fed. Any vertical rise or kink in drain line that is higher than chamber drain will result in a trap and not allow chamber to drain to recirculator. 8. Plug in power cord and turn on recirculator. INST-PM-CRSY102 Rev B 04/11/2018...