Temp Controller Failure
TMP CNTR ERR
Sensor Fault Alarms
4-2
Direct Heat CO2 Incubator
In addition to other safety features designed into Series 8000DH
incubators, a thermostat is provided to monitor the cabinet's temperature.
In the unlikely event of a temperature control failure, the thermostat will
turn off all heaters at a cabinet temperature of 160°C, ±5%. This is
intended to be a safety feature to protect the incubator, and is not intended
to protect the cell cultures or the equipment inside the chamber should a
temperature control failure occur. Should such a failure occur, contact the
Technical Services Department or your local distributor.
The microprocessor in Series 8000DH incubators continually scans all
available sensors to ensure that they are operating properly. Should an error
be detected, the incubator sounds an alarm and shows the appropriate
message. Contact the Technical Services Department (1-800-438-4851) or
your local distributor.