INSTRUCTIONS
Containment/Transportation
• Follow health care facility protocol for safe containment
and transport to the decontamination environment.
• It is recommended that devices are cleaned immediately
after use.
Preparation for Cleaning
• Disassembly is required.
Disassembling Equipment
Remove the gel pads from
the unit by sliding off. Keep
all joints in the unlocked
position. Slide the adjustable
headrest base until it comes
apart.
Cleaning – Manual Equipment: Water, Neutral pH
Detergent, Soft Bristle Brush, Towels
Method
1. Prepare neutral pH enzymatic detergent solution (e.g.
Endozime® AW Triple Plus with APA (Ruhof), 1:128 ratio)
according to detergent manufacturer's instructions
using lukewarm tap water.
2. Prepare equipment for soaking by disassembling
removable parts and loosening connections.
3. Rinse equipment in warm water before placing into
bath.
4. Completely soak equipment in water/detergent
solution for 30 minutes maximum.
5. Clean thoroughly with a soft nylon bristle brush.
NOTE: If possible, use a disposable brush.
6. Rinse in warm purified water until all visible substances
and residual detergent are removed. NOTE: Make sure
to give special attention to hard-to-reach areas.
7. Thoroughly dry equipment with soft clean towels
and use medically compressed air if needed, to dry
channels, crevices and lumens.
8. Inspect the equipment to make sure there is no visible
organic debris or residue from cleaning agent.
Repeat process if any soil is detected.
Drying
• Products should be dry at this point. If wetness or
excess liquid is detected, dry with a soft clean towel.
• Medically compressed air can be used if needed.
Inspecting for Cleanliness
Inspect the teeth of the
starbursts and make sure to
inspect all joints. Pay close
attention to the area around
the locking screw.
Optional Automatic Wash / Disinfect
CAUTIONS
• Alkaline and highly acidic detergents and solutions
cause damage to the devices.
• Channels and crevices found on these devices require
particular attention during cleaning.
• Pay special attention to the water quality used
throughout reprocessing. Hard water can damage
the surface of the equipment. Avoid using hard water.
Instead use purified water unless otherwise specified.
Limitations on Reprocessing
• Repeated processing has negative effects on these
devices and is not recommended for routine use.
• It is important to have Integra NeuroSpecialists perform
routine inspections (twice yearly is recommended).
See contact information.
INSTRUCTIONS
Cleaning - Automated Equipment: Neutral pH Detergent
Method
1. Prepare equipment for cleaning by disassembling
removable parts and loosening connections.
2. Rinse equipment in warm water before placing into
washer.
3. Load device into the washer and place small parts in
container or tray inside the washer unit in order to
avoid losing small components. NOTE: Load devices
carefully into washer in order to avoid collision.
4. Follow the instructions listed below and set washer
machine to these exact parameters:
Phase
Time
(Min.)
Pre-
4:00
wash 1
Enzyme
4:00
Wash
Wash 1
10:00
Rinse 1
0:30
Thermal
2:00
Rinse**
8
Water Tempera-
Detergent and
ture
Concentration
Cold water
N/A
Hot water
Neutral pH
enzymatic (e.g.
Endozime® AW
Triple Plus with
APA, Ratio 1:128)
60.0ºC
Neutral pH deter-
(140º F)
gent
(e.g. Renu-Klenz™,
Steris, Ratio 1:256)
Hot water
N/A
82.2ºC (180º F)
N/A