4.2.1 Wavelength Accuracy
[Summary]
To conduct: measuring emission lines from the D
486.0 nm), determining a wavelength having a maximum intensity,
and checking the deviation between the indicated wavelength and
the specified wavelength.
[Procedure]
‹General tab›
Measurement mode
Apply sample table
‹Instruemnt tab›
Data mode
Start wavelength (nm)
End wavelength (nm)
Scan speed (nm/min)
Response
Baseline correction
Initial waiting time (s)
AutoZero before measuring
Light source switching
Switching wavelength (nm)
Number of measuring times
Cell length (mm)
‹Monitor tab›
Open data processing window
after measurement
Open data processing window
after measurement with
spectrum superimposed
Automatic printing
on measuring end
Display spectrum on the screen
with superimposed
*1: On-choice setting is available. This setting does not
affect measuring.
*2: This item is masked for protection from a setting by an
operator. This setting does not affect measuring.
2. Choose [Edit] - [Sample name/Comment]. Enter, as the
sample name, "Wavelength accuracy"; then, enter the file
name.
3. Click the measuring button (
4. On completion of the measurement, open the data file of which
name was entered at the procedure 2.
4 - 5
lamp (656.1 nm,
2
Wavelength scan
OFF
E(R)
660/490
650/480
100
Fast
System (*1)
0
OFF
Use D
lamp only
2
340.0 (*2)
1
10 (*1)
ON
OFF
OFF
OFF
). (Measurement starts)
4.2.1