1.6 ANTIEARTHQUAKE MEASURE

Two threaded screw holes for a clamp are provided on both the
right and left sides of the body to secure the spectrophotometer
with an anti-earthquake measure.
Use these threaded screw holes to secure the photometer as
needed. (Each of the threaded screw holes is for a 4 mm screw,
and the spacing of two holes is 50 mm.) (The user needs to
prepare clamps and screws.)
Fig. 1-2. Example of securing spectrophotometer
Secure the personal computer on the laboratory table with retaining
bands as needed.
Fig. 1-3. Example of securing personal computer
1 - 5
Part No. 2J1-1365, Fastening Pla
Drawing of the clamp used in Fig.1-2
1.6
te
Stamping,
14 x 6 flat-oval hole,
2 holes
Stamping,
10 x 5 flat-oval hole,
2 holes