4.2 Performance Check
4.2.4 Noise Level
[Summary]
To determine a variation range of absorbance values measured
within a short time under a predetermined measurement conditions
with no sample in the sample compartment.
[Procedure]
1. Choose [Edit] - [Condition setting], then, set the conditions as
follows:
‹General tab›
Measurement mode
Apply sample table
‹Instruemnt tab›
Data mode
Measuring wavelength (nm)
Measuring time (s)
Response
Baseline correction
Initial waiting time (s)
AutoZero before measuring
Light source switching
Switching wavelength (nm)
Cell length (mm)
‹Monitor tab›
Horizontal axis maximum
Vertical axis minimum
Open data processing window
after measurement
Open data processing window
after measurement with
spectrum superimposed
Automatic printing
on measuring end
Display spectrum on the screen
with superimposed
*1: On-choice setting is available. This setting does not
affect measuring.
2. Choose [Edit] - [Sample name/Comment]. Enter, as the
sample name, "Noise level"; then, enter the file name.
3. Click the measuring button (
4. On completion of measurement, print the spectrum. If
needed, click the auto-scale button (
with the scale changed. Calculate difference between the
maximum and minimum (peak-to-peak) of the measured
photometrical values; then, divide the calculation by two
(multiply by 1/2).
4 - 10