On Cleaning
Before cleaning
Be sure to disconnect the AC power cord from the AC
outlet.
On cleaning the monitor
A material that withstands disinfection is used for the
front protection plate of the medical use LCD monitor.
The protection plate surface is specially treated to reduce
reflection of light. When solvents such as benzene or
thinner, or acid, alkaline or abrasive detergent, or
chemical cleaning cloth are used for the protection plate
surface/monitor surface, the performance of the
monitor may be impaired or the finish of the surface may
be damaged. Take care with respect to the following:
Clean the protection plate surface/monitor surface
with a 50 to 70 v/v% concentration of isopropyl
alcohol or a 76.9 to 81.4 v/v% concentration of ethanol
using a swab method. Wipe the protection plate
surface gently (wipe using less than 1 N force).
Stubborn stains may be removed with a soft cloth such
as a cleaning cloth lightly dampened with mild
detergent solution using a swab method and then
clean using the above chemical solution.
Never use solvents such as benzene or thinner, or acid,
alkaline or abrasive detergent, or chemical cleaning
cloth for cleaning or disinfection, as they will damage
the protection plate surface/monitor surface.
Do not use unnecessary force to rub the protection
plate surface/monitor surface with a stained cloth. The
protection plate surface/monitor surface may be
scratched.
Do not keep the protection plate surface/monitor
surface in contact with a rubber or vinyl resin product
for a long period of time. The finish of the surface may
deteriorate.
On Repacking
Do not throw away the carton and packing materials.
They make an ideal container which to transport the
unit.
If you have any questions about this unit, contact your
authorized Sony dealer.

Disposal of the Unit

Do not dispose of the unit with general waste.
Do not include the monitor with household waste.
For safety, do not connect the connector to peripheral
12
GB
Precaution
device wiring that might have excessive voltage.
Follow the instructions for use for the LAN port.
Do not come into contact with the terminals of the rear
panel connectors and patients at the same time.
Doing so may result in a generation of voltage that can be
harmful to patients if the unit is malfunctioning.
Always disconnect the power cord before connecting
and disconnecting connectors.
Using this unit for medical purposes
This equipment's connectors are not isolated.
Do not connect any device other than one which
conforms to IEC 60601-1.
When an information technology device or AV device
that uses an alternating current is connected, current
leakage may result in an electric shock to the patient or
operator.
If use of such a device is unavoidable, isolate its power
supply by connecting an isolation transformer, or by
connecting an isolator between the connecting cables.
After implementing these measures, confirm that the
reduced risk now conforms to IEC 60601-1.