On Fan Error; On Temperature Error; On Moisture Condensation; On Security - Sony LMD-X310MD Instructions For Use Manual

Table of Contents

Advertisement

Turn off the power if the monitor is not to be used for
a prolonged period of time.

On Fan Error

The fan for cooling the unit is built in. When the fan
error indication appears on the screen, turn off the
power and contact an authorized Sony dealer.

On Temperature Error

When this unit is used in a high temperature
environment and the internal temperature rises, an error
is displayed on the screen. When the error is displayed,
contact an authorized Sony dealer.

On Moisture Condensation

If the unit is suddenly taken from a cold to a warm
location, or if ambient temperature suddenly rises,
moisture may form on the outer surface of the unit and/
or inside of the unit. This is known as condensation. If
condensation occurs, turn off the unit and wait until the
condensation clears before operating the unit. Operating
the unit while condensation is present may damage the
unit.

On Security

SONY WILL NOT BE LIABLE FOR DAMAGES OF
ANY KIND RESULTING FROM A FAILURE TO
IMPLEMENT PROPER SECURITY MEASURES ON
TRANSMISSION DEVICES, UNAVOIDABLE DATA
LEAKS RESULTING FROM TRANSMISSION
SPECIFICATIONS, OR SECURITY PROBLEMS OF
ANY KIND.
Depending on the operating environment,
unauthorized third parties on the network may be able
to access the unit. When connecting the unit to the
network, be sure to confirm that the network is
protected securely.
This unit is equipped with a maintenance function
performed via a network. Maintenance may be
performed with your consent.

On Cleaning

Before cleaning
Be sure to disconnect the AC power cord from the AC
outlet.
14
Precaution
On cleaning the monitor
A material that withstands disinfection is used for the
front protection plate of the medical use LCD monitor.
The protection plate surface is specially treated to reduce
reflection of light. When solvents such as benzene or
thinner, or acid, alkaline or abrasive detergent, or
chemical cleaning cloth are used for the protection plate
surface/monitor surface, the performance of the
monitor may be impaired or the finish of the surface may
be damaged. Take care with respect to the following:
Clean the protection plate surface/monitor surface
with a 50 to 70 v/v% concentration of isopropyl
alcohol or a 76.9 to 81.4 v/v% concentration of ethanol
using a swab method. Wipe the protection plate
surface gently (wipe using less than 1 N force).
Stubborn stains may be removed with a soft cloth such
as a cleaning cloth lightly dampened with mild
detergent solution using a swab method and then
clean using the above chemical solution.
Never use solvents such as benzene or thinner, or acid,
alkaline or abrasive detergent, or chemical cleaning
cloth for cleaning or disinfection, as they will damage
the protection plate surface/monitor surface.
Do not use unnecessary force to rub the protection
plate surface/monitor surface with a stained cloth. The
protection plate surface/monitor surface may be
scratched.
Do not keep the protection plate surface/monitor
surface in contact with a rubber or vinyl resin product
for a long period of time. The finish of the surface may
deteriorate.

On Repacking

Do not throw away the carton and packing materials.
They make an ideal container which to transport the
unit.
If you have any questions about this unit, contact your
authorized Sony dealer.

Disposal of the Unit

Do not dispose of the unit with general waste.
Do not include the monitor with household waste.
About this manual
The instructions in this manual are for the following
models:
LMD-X310MD
LMD-X550MD
The illustration of LMD-X310MD is used for the
explanations. Any differences in specifications are
clearly indicated in the text.

Advertisement

Table of Contents
loading

This manual is also suitable for:

Lmd-x550md

Table of Contents