Immunity test
Proximity fields
from RF wireless
communica-
tions equipment
IEC 6100-4-3
10. H O T A I R G E N E R AT O R
IEC 60601
test level
380 - 390 MHz
27 V/m;
PM 50 %;
18 Hz
430 - 470 MHz
28 V/m;
(FM ±5 kHz,
1 kHz sine) PM;
18 Hz
704 - 787 MHz
9 V/m; PM 50 %;
217 Hz
800 - 960 MHz
28 V/m;
PM 50 %;
18 Hz
1700 - 1990 MHz
28 V/m;
PM 50 %;
217 Hz
2400 - 2570
MHz
28 V/m;
PM 50 %;
217 Hz
5100 - 5800
MHz
9 V/m; PM 50 %;
217 Hz
Compliance
level
380 - 390 MHz
27 V/m;
PM 50 %;
18 Hz
430 - 470 MHz
28 V/m;
(FM ±5 kHz,
1 kHz sine) PM;
18 Hz
704 - 787 MHz
9 V/m; PM 50 %;
217 Hz
800 - 960 MHz
28 V/m;
PM 50 %;
18 Hz
1700 - 1990 MHz
28 V/m;
PM 50 %;
217 Hz
2400 - 2570
MHz
28 V/m;
PM 50 %;
217 Hz
5100 - 5800
MHz
9 V/m; PM 50 %;
217 Hz
128
Electromagnetic
environment – Guidance
WARNING: Portable and
mobile RF communica-
tions devices (including
their accessories, such
as external antennas
and antenna cables)
should be used no closer
than 30 cm (12 inches)
to any part of the
VOLCANO MEDIC 2
Vaporization System,
including cables, speci-
fied by the manufacturer.
Otherwise, degradation
of the performance of
this equipment could
result.