9.1. General

If the VOLCANO MEDIC 2 Vaporization
System is to be used by other people
(patients), the Filling Chamber must
be cleaned and a new Valve Balloon
with Mouthpiece or a new Tube Kit
with Mouthpiece must be used. Fur-
ther information on the Filling Cham-
ber, Valve Balloon and Tube can be
found in Chapter 9.3."Filling Chamber
as well as Tube Flexer and Mouthpiece
of the Tube Kit" are described from
page 113 onwards.
The Filling Chamber, Valve Balloon
with Mouthpiece and Tube Kit with
Mouthpiece may only be used by one
person.
Caution!
To avoid a health risk, e.g. an
infection from a contaminat-
ed Filling Chamber, Valve Balloon or
Tube Kit, the following hygiene instruc-
tions must be followed.
The VOLCANO MEDIC 2 Vaporization
System is intended for multiple use.
a) Filling Chamber
The Filling Chamber must be
cleaned when a new patient uses it,
if there is visible residue or when a
new Valve Balloon is used.
b) Valve Balloon / Tube Kit
The Valve Balloon or Tube Kit may
only be used by one person.
Make sure that the equipment is
allowed
to
dry
9. H Y G I E N E
thoroughly
after
every
cleaning.
residual
moisture
age bacterial growth. Please note
Chapter
9.2.
ance of the VOLCANO MEDIC 2
Vaporization System", on this page .
Inspect the component parts of the
Filling Chamber regularly and replace
defective (broken, misshapen, dis-
colored) parts. Replace the Valve Bal-
loon or Tube Kit after no more than 70
applications or at least 14 days after
the first use.
9.2. Material Resistance of
Please make sure that the Filling
Chamber, Mouthpiece, Tube Flexer of
the Tube Kit and the Hot Air Generator
do not come into contact with the fol-
lowing chemicals:
organic, mineral and oxidating
acids (min. permissible pH 5.5)
strong bases (max. permissible pH
10). We recommend using neutral/
enzymatic or slightly alkaline clean-
ers.
organic solvents (e.g. aldehyde,
ether, ketone, benzene)
oxidation agents (e.g. hydrogen
peroxide)
halogens (chlorine, iodine, bro-
mine)
aromatic/halogenated hydrocar-
bons
112
Condensation
may
"Material
the VOLCANO MEDIC 2

Vaporization System

or