10.1. Cleaning

Before maintenance, the device has
to be switched off and the power plug
has to be unplugged.
For home use it is sufficient to clean
the housing of the Hot Air Generator
with a damp cloth. Do not hold the de-
vice under running water.
Alternatively, in domestic applications,
the Housing of the Hot Air Generator
can be cleaned and disinfected with a
surface disinfectant (details are given
in the following cleaning and disinfec-
tion instructions for use in doctor's
surgeries and hospitals).
In a physician's office or clinic, the
housing of the Hot Air Generator can
be cleaned and disinfected with a sur-
face disinfectant.
The surface disinfectant should be
certified (e.g. VAH/DGHM or FDA ap-
proval or CE marking), be suitable for
disinfecting the Hot Air Generator, and
compatible with the Hot Air Genera-
tor (see Chapter 9.2. "Material Resist-
ance", page 112).
Use only a soft clean cloth for wiping.
The cloth may be damp, but not wet.
Prevent liquid from entering the Hot
Air Generator.
The tests for the material resistance
of the Hot Air Generator for surface
disinfection were made using the dis-
infectant "neoform MED Spray" from
Dr. Weigert.
10. H O T A I R G E N E R AT O R
If a cleaning spray is being
used, it should be applied
onto a cloth with which the
device is then wiped. Do not spray
cleanser directly onto the device.
Liquid entering the Hot Air Generator
can damage the electrical parts and
cause malfunctions.
If any liquid does get into the Hot Air
Generator, contact our Service Center
immediately.

10.2. Care and Maintenance

Prior to undertaking any maintenance
work, the device must be switched
off and disconnected from the power
supply.
At the bottom of the Hot Air Generator
is an Air Filter.
The Air Filter must be inspected for
contamination every four weeks, and
changed as necessary. To do this,
open the rectangular flap on the bot-
tom of the Hot Air Generator and ex-
change the Air Filter.
Please pay attention to the cor-
rect positioning of the Air Fil-
ter: it has three large and one
small rounded corner.
119