9.4. Disassembly and Assembly of
the Filling Chamber
9.4.1. Filling Chamber and Filling
Chamber Cap
Please allow the Filling Chamber and
the Cap to cool down before disas-
sembling them.
All parts of the Filling Chamber must
be cleaned at the intervals prescribed
in Chapter 9.3. "Filling Chamber as well
as Tube Flexer and Mouthpiece of the
Tube Kit", page 113. Disassemble the
Filling Chamber as follows: Turn the
Filling Chamber Cap counterclockwise
in order to remove it from the Filling
Chamber.
Remove Filling Chamber Cap
9.4.2. Disassembly of the Filling
Chamber Cap
Use, e.g., the handle of the Cleaning
Brush to press the upper Screen down
in order to push it out. No Screens are
needed when using dronabinol.
9. H Y G I E N E
Push out the upper Screen with the help of
the Cleaning Brush
Strip off the cap ring. Push the cap
cylinder downwards out of the cap
housing.
Push the cap cylinder out of the cap housing
9.4.3. Disassembly of the Filling
Use Screen, or Drip Pad where applica-
ble, to push the Filling Chamber Cylin-
der upwards out of the Filling Chamber
Housing.
Push the Filling Chamber Cylinder out of the
Filling Chamber Housing
116